ToxSci Advance Access originally published online on December 1, 2005
Toxicological Sciences 2006 90(1):39-48; doi:10.1093/toxsci/kfj056
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Styrene-7,8-Oxide Burden in Ventilated, Perfused Lungs of Mice and Rats Exposed to Vaporous Styrene

,1
* Institute of Toxicology, GSF-National Research Center for Environment and Health, D-85764 Neuherberg, Germany; and
Institut für Toxikologie und Umwelthygiene, Technische Universität München, D-80802 München, Germany
1 To whom correspondence should be addressed at GSFNational Research Center for Environment and Health, Institute of Toxicology, Ingolstädter Landstrasse 1, D-85764 Neuherberg, Germany. Fax: +49 89 3187 3449. E-mail: johannes.filser{at}gsf.de.
Received July 1, 2005; accepted November 12, 2005
| ABSTRACT |
|---|
|
|
|---|
Styrene (ST) is an important industrial chemical. In long-term inhalation studies, ST-induced lung tumors in mice but not in rats. To test the hypothesis that the lung burden by the reactive metabolite styrene-7,8-oxide (SO) would be most relevant for the species-specific tumorigenicity, we investigated the SO burden in isolated lungs of male Sprague-Dawley rats and in-situ prepared lungs of male B6C3F1 mice ventilated with air containing vaporous ST and perfused with a modified Krebs-Henseleit buffer (37°C). Styrene vapor concentrations were determined in air samples collected in the immediate vicinity of the trachea. They were almost constant during each experiment. Styrene exposures ranged from 50 to 980 ppm (rats) and from 40 to 410 ppm (mice). SO was quantified from the effluent perfusate. Lungs of both species metabolized ST to SO. After a mathematical translation of the ex-vivo data to ventilation and perfusion conditions as they are occurring in vivo, a species comparison was carried out. At ST concentrations of up to 410 ppm, mean SO levels in mouse lungs ranged up to 0.45 nmol/g lung, about 2 times higher than in rat lungs at equal conditions of ST exposure. We conclude that the species difference in the SO lung burden is too small to consider the genotoxicity of SO as sufficient for explaining the fact that only mice developed lung tumors when exposed to ST. Another cause is considered as driving force for lung tumor development in the mouse.
Key Words: styrene; styrene-7,8-oxide; mouse; rat; ventilated, perfused lung; inhalation; mode of action; glutathione.
| INTRODUCTION |
|---|
|
|
|---|
Styrene (ST) is an important chemical of wide industrial use, particularly in the production of polymers, co-polymers, and reinforced plastics. Its toxicology and metabolic fate are described in several reviews, e.g., Cohen et al. (2002)
| MATERIAL AND METHODS |
|---|
|
|
|---|
Chemicals
All chemicals were of analytical grade if not indicated otherwise. Styrene (purity
99%), racemic styrene-7,8-oxide (purity 97%), (1R, 2R)-(+)-1-phenylpropenoxid (PPO; purity 97%), magnesium sulfate (99% anhydrous), albumin from bovine serum (purity
96%, essentially fatty acid free), HEPES buffer (minimum 99.5%, titration), and D-(+)-glucose (purity 99.5%) were obtained from Sigma-Aldrich, Steinheim, Germany. Acetonitrile (CHROMASOLV® for HPLC) was obtained from Riedel-de Haën, Seelze, Germany, isofluran from Baxter, Unterschleißheim, Germany, Ketamin (10%) from "Wirtschaftsgenossenschaft Deutscher Tierärzte eG," Garbsen, Germany, Rompun® (2%) from Bayer, Leverkusen, Germany, n-hexane (Picograde) from Promochem, Wesel, Germany, and Liquemin® N25000
[GenBank]
(Heparin-Natrium) from Hoffmann La Roche, Grenzach-Wyhlen, Germany. Gases used for gas chromatography were from Linde, Unterschleißheim, Germany. Diethyl maleate (purity 95%) and all other chemicals were obtained from Merck, Darmstadt, Germany.
Perfusion medium.
The perfusion medium of Uhlig and Heiny (1995)
was used: Krebs-Henseleit buffer (37°C) containing 2% bovine serum albumin, 0.1% glucose, and 0.3% HEPES. The pH was adjusted to 7.357.40.
Determination of ST.
Styrene was measured in air samples using a gas chromatograph (GC-8A, Shimadzu, Duisburg, Germany) equipped with a flame ionization detector (FID) and a stainless-steel column packed with Tenax TA, 6080 mesh, 1.5 m x 1/8''(Shimadzu, Duisburg). Air samples of 50 µl were injected directly on column. Separation was done isothermally at 200°C using N2 as carrier gas. Injector and detector temperatures were kept at 300°C. Retention time of ST was about 1.5 min. Chromatograms were recorded and integrated by a Shimadzu C-R5A integrator. Calibration curves were constructed three times by generating ST vapor concentrations ranging from 10 to 1000 ppm in atmospheres of closed desiccators according to the method of Filser et al. (1993)
. Calibration curves were linear in the whole range. Analysis of linear regression through the origin revealed correlation coefficients of at least 0.957 between peak areas and atmospheric styrene concentrations. Each time before starting an ST exposure, a one-point calibration was carried out in the concentration range used in the actual experiment.
Determination of SO.
Styrene oxide was quantified by gas chromatographyflame ionization detection (GC/FID) and additionally in at least one effluent perfusate sample per exposure experiment by gas chromatographymass selection detection (GC/MSD) for the unequivocal identification of SO. Fresh lung perfusate samples of 2.5 ml (rat) and 1 ml (mouse) were immediately spiked with a methanolic PPO solution of 0.5 µmol/l (rat 5 µl; mouse 1 µl) and mixed with acetonitrile (rat 2.5 ml; mouse 1 ml). n-Hexane (rat 4.5 ml; mouse 2 ml) was added and the mixture vigorously shaken for 2 min. Phase separation was done by centrifugation (5 min, 5°C, 4400 rpm). The clear supernatant was collected and the remaining aqueous phase extracted once again. The combined extracts were concentrated under a gentle stream of nitrogen to a final volume of 2.5 ml (rat) and 0.5 ml (mouse) and then stored at 4°C in a closed autosampler vial for a maximum time span of 2 days. GC/FID or GC/MSD analysis was carried out within this time frame. Under these conditions, SO and PPO were stable as had been demonstrated by Kessler et al. (1990)
.
The GC/FID determination of SO and PPO was done by capillary gas chromatography as described in Kessler et al. (1990)
. An Agilent GC 6090N equipped with a direct on-column injector was used. The retention times of SO (13.3 min) and of PPO (14.0 min) were very near to those reported in the study of Kessler et al. (1990)
. The limit of detection was about 30 nmol/l perfusate at an injection volume of 25 µl and a signal-to-noise ratio of 3:1.
The GC/MSD determination by capillary gas chromatography was done according to the method of Bitzenhofer (1993)
on a gas chromatograph HP 5890 series II equipped with a mass selective detector (HP 5970; both Agilent, Waldbronn, Germany. 1 µl of the concentrated n-hexane extract was manually injected onto a "fused silica" pre-column (length 10 m, ID 0.53 mm; from Agilent, Waldbronn, Germany) using the "Gerstel-KaltAufgabeSystem KAS 3" from Gerstel, Mühlheim an der Ruhr, Germany. The pre-column was connected to the separation column (HP-1 MS, film 0.33 µm, length 25 m, ID 0.2 mm; from Agilent, Waldbronn, Germany) by a "Miccon" press fit connector. Helium (0.8 ml/min) was used as carrier gas. Column temperature was maintained at 35°C for 0.25 min during the injection process. Then, it was heated up to 220°C with a rate of 30°C/min. The temperature of the transfer line to the MSD was 250°C. The electron ionization potential of the MSD was at 70 eV. Retention times were 7.7 min (SO) and 8.1 min (PPO). Both substances were detected in the single ion-monitoring mode at m/z 89 and 119 (SO) and 89 and 133 (PPO) and were quantified using the m/z 89 according to Langvardt and Nolan (1991)
. The detection limit for SO was about 15 nmol/l perfusate at an injection volume of 1 µl and a signal-to-noise ratio of 3:1.
Calibration curves (signal area of SO to signal area of PPO versus the concentration of SO) were constructed over a concentration range from 0.1 µmol/l to 10 µmol/l using five different SO concentrations. Analysis of linear regression through the origin revealed correlation coefficients of at least 0.999 (both GC/FID and GC/MS) between peak areas and SO concentrations. At each ST perfusion experiment, a three-point calibration curve through the origin was constructed.
Animals and Surgery
Male Spague-Dawley rats and male B6C3F1 mice were obtained from Charles River Wiga GmbH, Sulzfeld, Germany. All experimental procedures with animals were performed in conformity with the "Guide for the Care and Use of Laboratory Animals" (7th edition, Institute of Laboratory Animal Resources, Commission on Life Sciences, National Research Council, National Academy Press Washington, D.C., 1996) under the surveillance of the authorized representative for animal welfare of GSF. Up to 4 weeks before use, two rats or five mice each were housed in the GSF-Institute of Toxicology in a Macrolon type III cage placed in an IVC top-flow system (Tecniplast, Buguggiate, Italy). This system provided the animals with filtered room air. A constant 12-h light/dark cycle was maintained in the chamber room. Animals had free access to standard chow (Nr. 1324 from Altromin; Lage, Germany) and tap water. When using the animals, body weights were in the range of 230360 g (rats) and 2534 g (mice).
Lungs were prepared as described in Uhlig and Wollin (1994)
for rats and in von Bethmann et al. (1998)
for mice with the exceptions that different narcotics were used instead of pentobarbital sodium. To rats, a mixture of 0.8 ml/kg ketamine and 1 ml/kg Rompun was administered. Per kg body weight, mice received a mixture of 1.4 ml ketamine, 0.4 ml Rompun, and 6.3 ml of a 0.9% saline. Immediately before intraperitoneal injection of this mixture, animals were anesthetized by isoflurane inhalation.
In summary, animals were intubated and ventilated by positive pressure (rats: 80 breath/min, tidal volume approximately 2 ml; mice: 90 breath/min, tidal volume approximately 200 µl) using the ventilationperfusion systems for rat and mouse lungs "Isolierte perfundierte Lunge Größe 2" (rat) and "Isolierte perfundierte Lunge Größe 1" (mouse) manufactured by Hugo Sachs Elektronik (HSE), March-Hugstetten, Germany. After laparotomy, they received intracardial injections of Liquemin® N25000 [GenBank] (0.1 ml per rat and 0.05 ml per mouse). Thereafter, the diaphragms were removed and the animals were exsanguinated. A ligature was placed around both the pulmonary artery and the aorta. The cannula for the influent perfusate was inserted into the pulmonary artery and fixed by the ligature thereby closing the aorta. Immediately, the apex cordis was opened by a cut. Then the cannula for the effluent perfusate was inserted into the left ventricle and also fixed by a ligature. The rat lung together with the heart were carefully isolated and suspended by the trachea in the humidified and warmed rat lung ventilation chamber. The lid was closed and the lung was immediately ventilated using negative pressure. The mouse lung was prepared in the open mouse ventilation chamber and was not separated from the body. After removing the thorax, the chamber lid was closed and negative pressure ventilation was started.
Experimental Procedure
For safety reasons, all exposure experiments were carried out under the hood. The once-through perfused lungs were ventilated by negative pressure with ST vapors of defined concentrations in the inhaled air, and SO was measured in the effluent perfusate. Thereafter, the perfusion and ventilation parameters were extrapolated to the species-specific in vivo condition, and the corresponding SO lung burdens were calculated by means of the partition coefficient lung-to-perfusate of SO.
Ventilationperfusion of rat and mouse lungs.
Both above-described exposure systems from HSE, described in Uhlig and Wollin (1994)
and Uhlig and Heiny (1995)
for the rat as well as in von Bethmann et al. (1998)
for the mouse, were modified to enable inhalation exposure to a gas of constant concentration.
Modifications of the rat system for inhalation exposure.
Figure 1 presents a scheme of the mechanical parts of the isolated perfused lung system for rats including the following modifications: A T-formed hollow tube (T-tube) made in-house out of brass was cold-welded to the upper end of the pneumotachograph tube and connected via a Tygon® tubing to the ST-vaportransporting glass pipe. A constant flow of ST vapor from the "ST-stock desiccator" through the glass pipe and the horizontal duct of the T-tube was established (flow of ST vapor containing air: 15 ml/min) by means of the roller pump I that contained Tygon tubing. The lung took its breathing air from the T-tube via the pneumotachograph tube. In order to enable the determination of ST vapor concentrations immediately in front of the trachea, a custom-made "Kel-F" hub needle (Hamilton, Bonaduz, Switzerland) was directed through a small pipe in the lid of the artificial thorax into a lateral outlet of the trachea attachment and affixed gas-tight by polytetrafluoroethylene (PTFE; Teflon) and silicone tubing. The upper end of this cannula possessed a Teflon luer lock usually closed by a plug (item 5, Fig. 1). For measuring ST concentrations in inhaled air, the plug was exchanged by a Hamilton syringe (100 µl series 1710 TLL, obtained from Machery-Nagel, Düren, Germany), and a gas sample of 50 µl was collected and immediately injected on the column of the GC-8A.
|
Modifications of the mouse system for inhalation exposure.
Figure 2 presents a scheme of the mechanical parts of the isolated perfused lung system for mice, including the modifications enabling inhalation exposure. The original system is equipped with a moistener for inhaled air (item 11, Fig. 2). This vial (volume about 2 cm3), which is covered by a PTFE plug, contains two openings, one to the external air and the other to the pneumotachograph (item 8, Fig. 2). After drilling a hole through the PTFE plug, a small PTFE hose was inserted tightly. The ST vapor was transferred from the ST-stock desiccator by the roller pump I (11.3 ml/min) into the air space of the moistener. The ventilated lung inhaled the ST vapor from the air mixture passing through the modified moistener. To determine the actual inhaled ST concentration, the 51-mm-long needle of a Hamilton syringe (100 µl, series 1710 TLL) was inserted into the trachea through the opening of the moistener to the room air (item 5, Fig. 2), and an air sample of 50 µl was collected and immediately injected on the column of the GC-8A.
|
Exposure conditions.
Rat lungs (n = 13) were ventilated in the negative chamber pressure mode as described in Uhlig and Wollin (1994)
|
In all mouse experiments (n = 12), lungs were ventilated in the negative chamber pressure mode and were perfused as described in von Bethmann et al. (1998)
|
The values of these parameters were controlled and monitored by a computer running the software ( "Pulmodyn W" from HSE) of the ventilationperfusion system that possessed the "PLUGSYS" unit (from HSE) to amplify the raw data.
To test for the integrity of the alveolar membranes, lactate dehydrogenase (LDH) release into the perfusate was measured using the test kit "LDH IFCC-Methode 37°C" (Rolf Greiner BioChemica, Flacht, Germany). Mean activity (±S.D.) in the influent perfusates (background controls) was 7.1 ± 4.3 U/l (n = 25). In the effluent perfusates it was 6.7 ± 4.3 U/l (n = 93). The maximum values in the influent and effluent perfusate were identical: 17.6 U/l. Consequently, the lungs were considered to be physiologically and biochemically intact.
Partition coefficient lung-to-perfusate of SO (PL/P).
The thermodynamic partition coefficient PL/P depends on the temperature and on physicochemical properties of SO, lung tissue, and perfusate. For tissue and perfusate, water and lipid contents are most relevant. PL/P had to be determined (at 37°C) because it enabled calculation of the actual SO concentration in the cellular tissue (CLSOc) of the ST-inhaling lung at steady state from measurements of SO in the vasculature (the effluent perfusate; CPSO). Under this condition, CLSOc is in thermodynamic equilibrium with the perfusate and can therefore be calculated by the relationship:
![]() | (1) |
About 30 min before preparation of the rat lungs for determination of PL/P, the animals were administered intraperitoneally diethyl maleate (0.6 ml/kg body weight i.p.), a glutathione-depleting compound (see Plummer et al., 1981
). Thereafter, rats were narcotized and the lungs were positively ventilated and perfused in situ (see Animals and Surgery). Perfusion was carried out with the above-described perfusion medium for 10 min with a flow of 20 ml/min in order to remove blood completely from the lung. Because only one ventilation-and-perfusion system was available and three rat lungs were required for one experiment, the lungs had to be stored before use. Therefore, they were flash frozen in liquid nitrogen and stored in a freezer at 80°C.
After cutting the freshly thawed lungs into halves, each half was placed in a glass beaker (10 ml) together with enough physiological saline to cover it completely. Then, the beakers were placed for 5 min in a boiling water bath to denature the SO-metabolizing enzymes. Thereafter, lung-halves were taken from the beakers, carefully dabbed with a tissue, weighed, and cut in small slices. Each sliced lung-half was placed in a glass culture tube (7 ml; with a PTFE covered screw cab; K&K Laborbedarf, München, Germany) to which 1 ml of perfusion medium was added to a defined SO concentration (1 or 10 µmol/l). The tubes were sealed and immediately incubated in a shaking water bath (37°C). Parallel SO incubations were carried out in perfusion medium alone to determine loss of SO by non-enzymic hydrolysis. Aliquots of 0.5 ml perfusion medium (one vial containing one lung-half represented one time point) were taken at selected time points, PPO (5 µmol/l) was added and the SO content in the liquid phase was determined by GC/MSD as described above. From the concentration-time courses of SO obtained in the incubations with or without lung, the PL/P was calculated by means of a two-compartment model describing distribution between compartment one (medium) and compartment two (lung) and elimination from compartment one by hydrolysis. The partition coefficient was calculated using the following equation:
![]() | (2) |
- k1, k2 = micro-constants describing the SO transport between both compartments
- V1 = volume of compartment one (medium)
- V2 = volume of compartment two (lung)
- V1 = volume of compartment one (medium)
![]() | (3) |
- cSO = concentration of SO in medium at time point t;
- C1, C2 = intercepts of the function;
, ß = rate constants of the function.
- C1, C2 = intercepts of the function;
![]() | (4a) |
![]() | (4b) |
![]() | (5) |
- cSO = SO concentration in the medium without lung at any time point t;
- C(0) = SO concentration in the medium at time point t = 0;
- kel = elimination rate constant.
- C(0) = SO concentration in the medium at time point t = 0;
, and ß. These values were inserted in Equation 4 to calculate k1 and k2. PL/P was then calculated using Equation 2.
Extrapolation from ex vivo to in vivo conditions.
The perfusion flows used ex vivo in the ventilated perfused lungs were smaller than those occurring in vivo. This reduction was inevitable to avoid edema formation. Estimation of the in vivo burden of the lungs is required for a species comparison of the SO lung burden between mice and rats that results from exposure to ST. To calculate the in vivo burden resulting only from ST metabolism in the lung (i.e., not considering the lung burden by SO resulting from ST metabolism in the liver), the following extrapolation from the ex vivo to the in vivo situation was done considering the relationship expressed by Equation 1:
![]() | (6) |
- PL/P = partition coefficient lung/perfusate of SO*;
- CSOLiv = SO concentration in lung in vivo [nmol/g];
- CSOP = SO concentration in effluent perfusate (ex vivo) [nmol/ml];
- Qaiv = alveolar ventilation in vivo (mouse: 25; rat: 117 [ml/min])**;
- Qaev = actual alveolar ventilation ex vivo [ml/min] (calculated as actual tidal volume times fixed breath frequency times 2/3***);
- QLev = lung perfusion ex vivo [ml/min];
- QLiv = lung perfusion in vivo (mouse: 17; rat: 83 [ml/min])**;
- * = determined at 37°C;
- ** = from Arms and Travis (1988)
;
- *** = The number "2/3" represents the conversion factor from pulmonary to alveolar ventilation (e.g., Arms and Travis 1988
).
- CSOLiv = SO concentration in lung in vivo [nmol/g];
Calculations and Statistics
Means ± standard deviations (SD) of measured data, comparison of means by the unpaired two-tailed t-test, as well as the values of kinetic parameters describing measured concentrationtime courses were calculated using Prism 4 for Macintosh (GraphPad Software, San Diego, CA, USA).
| RESULTS |
|---|
|
|
|---|
Partition Coefficient Lung-to-Perfusate of SO (PL/P)
Partition coefficient lung-to-perfusate (PL/P) was calculated using the parameters gained by curve fitting from 7 concentrationtime courses of SO in perfusate that were recorded up to 250 min at pH 7.4 and 37°C (between 5 and 6 data points per concentrationtime course). Four of these experiments were conducted in SO containing perfusate free of lung tissue and the other three with heat-inactivated lung tissue in the perfusate at initial SO concentrations between 0.55 an 10 µmol/l (Fig. 3). For these experiments, the kinetic parameters describing the fitted curves, the rate constants of the spontaneous hydrolysis (kel), and the calculated PL/P values are given in Table 1. The half-life (ln2/kel) of the spontaneous hydrolysis at pH 7.4 was (mean ± SD; n = 4) 298 ± 27 min. This value fits to that of 306 min determined earlier in potassium phosphate buffer at pH 7.0 (Filser et al., 1992
|
|
SO Burden in ST Inhaling Lungs
Figures 4 and 5 depict concentrationtime courses that are characteristic for the inhalation studies with ventilated, perfused lungs of rats (Fig. 4) and mice (Fig. 5). Atmospheric ST concentrations in the air entering the lungs were maintained nearly constant during the exposure time (Figs. 4A and 5A). Concentrations of metabolically produced SO in the effluent perfusate remained nearly constant from about 5 min after starting ST exposures until the end of the experiments (Figs. 4B and 5B). In both figures, the straight lines represent the mean concentrations of the measured data from about 5 min after starting the respective exposure to its end.
|
|
The experimentally determined mean SO concentrations in effluent perfusate (Tables 2 and 3) were mathematically translated from the experimental to the species-specific in vivo conditions of lung perfusion and ventilation (see Eq. 6). These translated ( "measured") lung levels of SO are depicted in Figure 6 (both species) versus the corresponding ST concentrations in inhaled air. Over the whole demonstrated concentration range of ST, the "measured" pulmonary SO burden is about 2 times lower in rat than in mouse. At ST concentrations of 100 ppm and above, the SO levels in rat lungs are higher than in mouse lungs at 40 ppm. Additionally, the SO levels in rat lungs at 350 ppm ST and above are similar to those in mouse lungs at 160 ppm ST.
|
| DISCUSSION |
|---|
|
|
|---|
The pulmonary SO levels resulting solely from ST metabolism in the lung (Fig. 6), are at least 5 times (mice) and 10 times smaller than the SO blood levels that were reached during inhalation exposure of rats and mice to equal ST concentrations of up to 250 ppm. From the blood data in rats and mice measured (Kessler et al., 1992
Considering that in a long-term inhalation study (Cruzan et al., 2001
) with mice exposed to concentrations of up to 160 ppm the frequency of lung tumors had increased at ST concentrations
20 ppm, and that no tumors had been found in lungs of rats exposed up to 1000 ppm ST under comparative conditions (Cruzan et al., 1998
), one has to conclude that the species-specific lung tumorigenicity of ST in mice does not result directly from the lung burden by the genotoxic metabolite SO. As already concluded from the results delivered by our ST model, it is highly probable that a non-genotoxic glutathione-dependent mechanism is involved (Filser et al., 2002
).
Based on modeled tissue concentrations of SO and on kinetic interactions of the glutathione turnover with the SO forming and eliminating enzyme activities that had been measured in vitro (Oberste-Frielinghaus et al., 1999
), our physiological toxicokinetic model (Csanády et al., 2003
) predicted a drastic glutathione decrease of 40% in the bronchiolar region of mice and only a marginal one in that of rats following a 6-h inhalation exposure to 20 ppm ST. Although model predictions for the whole lung agreed with measured data (Dhawan-Robl et al., 2000
; Filser et al., 2002
), it cannot be ruled out that other ST metabolites might also contribute to the glutathione depletion in mouse lungs. For instance, 4-vinylphenole, a putative ST metabolite in mice (Carlson et al., 2001
) and a minor one detected in urine of rats (Bakke and Scheline, 1970
; Manini et al., 2002
; Pantarotto et al., 1978
) and humans (Manini et al., 2002
; Pfäffli et al., 1981
), resulted in a certain glutathione depletion in the lungs of mice, following intraperitoneal administration (Turner et al., 2005
). 4-Vinylphenol was hepato- and pneumotoxic in rodents due to the formation of metabolites, probably ring-opened products (Carlson et al., 2001
; Carlson 2002
).
Repeated administration of mouse Clara cell toxicants as coumarin, naphthalene, or 4-ipomeanol can result in a certain tolerance to these chemicals, probably as a result of upregulation of the detoxifying glutathione (Boyd et al., 1981
; Born et al., 1999
; Vassallo et al., 2004
; West et al., 2000
). However, after intermittence of daily injections of naphthalene for 4 days, elevated glutathione levels in the terminal airways of tolerant mice had declined to control levels and mice were again susceptible to naphthalene injury (West et al., 2000
). Plopper et al. (2001)
have investigated early events in naphthalene-induced acute Clara cell toxicity. They described a broad heterogenicity of loss of cellular glutathione and other sulfhydryls in the cell populations of the minimally susceptible lobar and proximal bronchi. However, in the most susceptible distal bronchioles, high loss of glutathione was detected in all cells. The authors concluded that at least 50% of the intracellular glutathione pool must be lost before cell organelle changes become apparent and that a loss of at least 75% is required before toxic cellular changes become irreversible. The observed cytotoxic effects should result at least directly from the glutathione depletion and not from a reactive naphthalene metabolite, because Phimister et al. (2005)
demonstrated that abrupt glutathione depletion in mice resulted in Clara cell toxicity similar to naphthalene treatment with respect to Clara cell swelling, plasma membrane blebs, and actin cytoskeleton disruptions.
Furthermore, it is well known that severe changes in the glutathione homeostasis can lead to apoptosis and cell proliferation (reviewed in, e.g., Rahman et al., 1999
). Together with a depletion of glutathione in lung homogenate, such effects have been detected in Clara cells of mice repeatedly exposed to ST (Gamer et al., 2004
). Following repeated SO or 4-vinylphenol administration to mice, cell proliferation, histomorphological changes, and apoptosis in bronchi and terminal bronchioles were observed. It was suggested that Clara cells were primary target cells (Kaufman et al., 2005).
Considering this information, we propose the following hypothesis to explain lung tumor formation in ST exposed mice: glutathione conjugation with the ST metabolite SO results in perturbation of the glutathione pathway in Clara cells of the terminal bronchioles because of insufficient glutathione turnover. This effect leads to cell death and regenerative proliferation. Together with the cell burdens by reactive ST metabolites as SO and possibly ring-oxidized derivatives, the finally observed tumorigenesis in mouse lungs becomes comprehensible. In the rat lung, no such effects are to be expected, because ST-induced loss of glutathione is minimal in this species (see Filser et al., 2002
; Csanády et al., 2003
).
| NOTES |
|---|
This publication is dedicated to Prof. Dr. Helmut Greim on the occasion of his 70th birthday.
| ACKNOWLEDGMENTS |
|---|
We thank Prof. Dr. Albrecht Wendel, Biochemical Pharmacology, University of Konstanz, Germany, and Prof. Dr. Stefan Uhlig, Division of Pulmonary Pharmacology, Research Center Borstel, Leibniz-Center for Medicine and Biosciences, Borstel, Germany, for friendly advice concerning the perfused, ventilated lung system. Financial support by the Styrene Steering Committee (SSC) of the European Chemical Industry Council (CEFIC) is gratefully acknowledged.
| REFERENCES |
|---|
|
|
|---|
Arms, A. D., and Travis, C. C. (1988). Reference physiological parameters in pharmacokinetic modeling. EPA/600/688/004, US Environmental Protection Agency.
Bakke, O. M., and Scheline, R. R. (1970). Hydroxylation of aromatic hydrocarbons in the rat. Toxicol. Appl. Pharmacol. 16, 691700.[CrossRef][ISI][Medline]
Bitzenhofer, U. N. (1993). Entwicklung einer gasdichten Apparatur zur Untersuchung von flüchtigen Fremdstoffen mit der isoliert perfundierten Rattenleber Kinetik von Styrol und seinem Metaboliten Styrol-7,8-oxid. GSF-Bericht 24/93, GSFForschungszentrum für Umwelt und Gesundheit, GmbH (Editor), Ingolstädter Landstraße 1, D-85764 Neuherberg, Germany, ISSN 0721-1694.
Boogaard, P. J., de Kloe, K. P., Wong, B. A., Sumner, S. C., Watson, W. P., and van Sittert, N. J. (2000). Quantification of DNA adducts formed in liver, lungs, and isolated lung cells of rats and mice exposed to (14)C-styrene by nose-only inhalation. Toxicol. Sci. 57, 203216.
Born, S. L., Fix, A. S., Caudill, D., and Lehman-McKeeman, L. D. (1999). Development of tolerance to Clara cell necrosis with repeat administration of coumarin. Toxicol. Sci. 51, 300309.
Boyd, M. R., Burka, L. T., Wilson, B. J., and Sastry, B. V. (1981). Development of tolerance to the pulmonary toxin, 4-ipomeanol. Toxicology 19, 85100.[CrossRef][ISI][Medline]
Cantoni, L., Salmona, M., Facchinetti, T., Pantarotto, C., and Belvedere, G. (1978). Hepatic and extrahepatic formation and hydratation of styrene oxide in vitro in animals of different species and sex. Toxicol. Lett. 2, 179186.[CrossRef]
Carlson, G. P. (1997). Effects of inducers and inhibitors on the microsomal metabolism of styrene to styrene oxide in mice. J. Toxicol. Environ. Health 51, 477488.[CrossRef][ISI][Medline]
Carlson, G. P., Perez Rivera, A. A., and Mantick, N. A. (2001). Metabolism of the styrene metabolite 4-vinylphenol by rat and mouse liver and lung. J. Toxicol. Environ. Health A 63, 541551.[CrossRef][ISI][Medline]
Carlson, G. P. (2002). Effect of the inhibition of the metabolism of 4-vinylphenol on its hepatotoxicity and pneumotoxicity in rats and mice. Toxicology 179, 129136.[CrossRef][ISI][Medline]
Cohen, J. T., Carlson, G., Charnley, G., Coggon, D., Delzell, E., Graham, J. D., Greim, H., Krewski, D., Medinsky, M., Monson, R., et al. (2002). A comprehensive evaluation of the potential health risks associated with occupational and environmental exposure to styrene. J. Toxicol. Environ. Health B Crit. Rev. 5, 1265.[CrossRef][ISI][Medline]
Conti, B., Maltoni, C., Perino, G., and Ciliberti, A. (1988). Long-term carcinogenicity bioassays on styrene administered by inhalation, ingestion and injection and styrene oxide administered by ingestion in SpragueDawley rats, and para-methylstyrene administered by ingestion in SpragueDawley rats and mice. Ann. N. Y. Acad. Sci. 534, 203234.[Abstract]
Cruzan, G., Cushman, J. R., Andrews, L. S., Granville, G. C., Johnson, K. A., Hardy, C. J., Coombs, D. W., Mullins, P. A., and Brown, W. R. (1998). Chronic toxicity/oncogenicity study of styrene in CD rats by inhalation exposure for 104 weeks. Toxicol. Sci. 46, 266281.
Cruzan, G., Cushman, J. R., Andrews, L. S., Granville, G. C., Johnson, K. A., Bevan, C., Hardy, C. J., Coombs, D. W., Mullins, P. A., and Brown, W. R. (2001). Chronic toxicity/oncogenicity study of styrene in CD-1 mice by inhalation exposure for 104 weeks. J. Appl. Toxicol. 21, 185198.[CrossRef][ISI][Medline]
Csanády, G. A., Mendrala, A. L., Nolan, R. J., and Filser, J. G. (1994). A physiologic pharmacokinetic model for styrene and styrene-7,8-oxide in mouse, rat and man. Arch. Toxicol. 68, 143157.[CrossRef][ISI][Medline]
Csanády, G. A., Kessler, W., Hoffmann, H. D., and Filser, J. G. (2003). A toxicokinetic model for styrene and its metabolite styrene-7,8-oxide in mouse, rat and human with special emphasis on the lung. Toxicol. Lett. 138, 75102.[CrossRef][ISI][Medline]
Dhawan-Robl, M., Oberste-Frielinghaus, H., Csanády, Gy. A., and Filser, J. G. (2000). Depletion of pulmonary glutathione in mouse and rat caused by inhalation of styrene. Naunyn-Schmiedeberg's Arch. Pharmacol. Suppl. 361, R139.
Filser, J. G., Kessler, W., Schwegler, U., Greim, H., and Hoffmann, H. D. (1992). Studies on toxicokinetics and macromolecular binding of styrene. Vol 1. Studies on the kinetics of styrene and styrene oxide in rats and mice. ECETOC, Special Report No. 3.
Filser, J. G., Schwegler, U., Csanády, Gy. A., Greim, H., Kreuzer, P. E., and Kessler, W. (1993). Species-specific pharmacokinetics of styrene in rat and mouse. Arch. Toxicol. 67, 517530.[CrossRef][ISI][Medline]
Filser, J. G., Kessler, W., and Csanády, Gy. A. (2002). Estimation of a possible tumorigenic risk of styrene from daily intake via food and ambient air. Toxicol. Lett. 126, 118.[CrossRef][ISI][Medline]
Gamer, A. O., Leibold, E., Deckardt, K., Kittel, B., Kaufmann, W., Tennekes, H. A., and van Ravenzwaay, B. (2004). The effects of styrene on lung cells in female mice and rats. Food Chem. Toxicol. 42, 16551667.[CrossRef][ISI][Medline]
Gram, T. E. (1997). Chemically reactive intermediates and pulmonary xenobiotic toxicity. Pharmacol. Rev. 49, 297341.
Hynes, D. E., DeNicola, D. B., and Carlson, G. P. (1999). Metabolism of styrene by mouse and rat isolated lung cells. Toxicol. Sci. 51, 195201.
IARC (1994). Monographs on the evaluation of carcinogenic risks to humans. Some industrial chemicals, Volume 60, WHO, International Agency for Research on Cancer, Lyon, France.
IARC (2002). Monographs on the evaluation of carcinogenic risks to humans. Some traditional herbal medicines, some mycotoxins, naphthalene and styrene, Volume 82, WHO, International Agency for Research on Cancer, Lyon, France.
Kaufmann, W., Mellert, W., van Ravenzwaay, B., Landsiedel, R., and Poole, A. (2005). Effects of styrene and its metabolites on different lung compartments of the mousecell proliferation and histomorphology. Regul. Toxicol. Pharmacol. 42, 2436.[CrossRef][ISI][Medline]
Kessler, W., Jiang, X., and Filser, J. G. (1990). Direct determination of styrene-7,8-oxide in blood by gas chromatography with flame ionization detection. J. Chromatogr. 534, 6775.[Medline]
Kessler, W., Jiang, X., and Filser, J. G. (1992). Pharmacokinetik von Styrol-7,8-oxid bei Maus und Ratte. In: Kreutz R. & Piekarski C., eds, 32nd Annual Meeting of the German Society of Occupational Medicine, Köln, 1992, Stuttgart, Gentner Verlag, pp. 622626.
Langvardt, P. W., and Nolan, R. J. (1991). Determination of styrene-7,8-oxide in whole blood by gas chromatography-mass spectrometry. J. Chromatogr. 567, 93103.[Medline]
Lijinsky, W. (1986). Rat and mouse forestomach tumors induced by chronic oral administration of styrene oxide. J. Natl. Cancer Inst. 77, 471476.[ISI][Medline]
Löf, A., Gullstrand, E., Lundgren, E., and Nordqvist, M. B. (1984). Occurrence of styrene-7,8-oxide and styrene glycol in mouse after the administration of styrene. Scand. J. Work Environ. Health 10, 179187.[ISI][Medline]
Manini, P., Andreoli, R., Poli, D., De Palma, G., Mutti, A., and Niessen, W. M. (2002). Liquid chromatography/electrospray tandem mass spectrometry characterization of styrene metabolism in man and in rat. Rapid. Commun. Mass Spectrom. 16, 22392248.[CrossRef][Medline]
Mendrala, A. L., Langvardt, P. W., Nitschke, K. D., Quast, J. F., and Nolan, R. J. (1993). In vitro kinetics of styrene and styrene oxide metabolism in rat, mouse, and human. Arch. Toxicol. 67, 1827.[CrossRef][ISI][Medline]
Morgan, D. L., Mahler, J. F., Dill, J. A., Price, H. C., Jr., O'Connor, R. W., and Adkins, B. Jr. (1993). Styrene inhalation toxicity studies in mice. II. Sex differences in susceptibility of B6C3F1 mice. Fundam. Appl. Toxicol. 21, 317325.[CrossRef][ISI][Medline]
Nakajima, T., Wang, R. S., Elovaara, E., Gonzalez, F. J., Gelboin, H. V., Vainio, H., and Aoyama, T. (1994). CYP2C11 and CYP2B1 are major cytochrome P450 forms involved in styrene oxidation in liver and lung microsomes from untreated rats, respectively. Biochem. Pharmacol. 48, 637642.[CrossRef][ISI][Medline]
Oberste-Frielinghaus, H., Dhawan-Robl, M., Pütz, C., Csanády, Gy. A., Baur, C., and Filser, J. G. (1999). Metabolism of styrene and racemic styrene-7,8-oxide in microsomes and cytosol of lung obtained from mouse, rat and human. Naunyn-Schmiedeberg's Arch. Pharmacol. Suppl. 359, R157.
Pantarotto, C., Fanelli, R., Bidoli, F., Morazzoni, P., Salmona, M., and Szczawinska, K. (1978). Arene oxides in styrene metabolism, a new perspective in styrene toxicity? Scand. J. Work Environ. Health 4(Suppl. 2), 6777.
Pfäffli, P., Hesso, A., Vainio, H., and Hyvonen, M. (1981). 4-Vinylphenol excretion suggestive of arene oxide formation in workers occupationally exposed to styrene. Toxicol. Appl. Pharmacol. 60, 8590.[CrossRef][ISI][Medline]
Phimister, A. J., Williams, K. J., Van Winkle, L. S., and Plopper, C. G. (2005). Consequences of abrupt glutathione depletion in murine Clara cells: Ultrastructural and biochemical investigations into the role of glutathione loss in naphthalene cytotoxicity. J. Pharmacol. Exp. Ther. 314, 506513.
Plummer, J. L., Smith, B. R., Sies, H., and Bend, J. R. (1981). Chemical depletion of glutathione in vivo. Methods Enzymol. 77, 5059.[Medline]
Plopper, C. G., Van Winkle, L. S., Fanucchi, M. V., Malburg, S. R. C., Nishio, S. J., Chang, A., and Buckpitt, A. R. (2001). Early events in naphthalene-induced acute clara cell toxicity II. Comparison of glutathione depletion and histopathology by airway location. Am. J. Respir. Cell Mol. Biol. 24, 272281.
Ponomarkov, V., Cabral, J., Wahrendorf, J., and Galendo, D. (1984). A carcinogenicity study of styrene-7,8-oxide in rats. Cancer Lett. 24, 95101.[CrossRef][ISI][Medline]
Rahman, Q., Abidi, P., Afaq, F., Schiffmann, D., Mossman, B. T., Kamp, D. W., and Athar, M. (1999). Glutathione redox system in oxidative lung injury. Crit. Rev. Toxicol. 29, 543568.[CrossRef][ISI][Medline]
Salmona, M., Pachecka, J., Cantoni, L., Belvedere, G., Mussini, E., and Garattini, S. (1976). Microsomal styrene mono-oxygenase and styrene epoxide hydrase activities in rats. Xenobiotica 6, 585591.[ISI][Medline]
Sarangapani, R., Teeguarden, J. G., Cruzan, G., Clewell, H. J., and Andersen, M. E. (2002). Physiologically based pharmacokinetic modeling of styrene and styrene oxide respiratory-tract dosimetry in rodents and humans. Inhalat. Toxicol. 14, 789834.[CrossRef][ISI][Medline]
Sumner, S. J., and Fennell, T. R. (1994). Review of the metabolic fate of styrene. Crit. Rev. Toxicol. 24(Suppl.), S11S33.
Turner, M., Mantick, N. A., and Carlson, G. P. (2005). Comparison of the depletion of glutathione in mouse liver and lung following administration of styrene and its metabolites styrene oxide and 4-vinylphenol. Toxicology 206, 383388.[CrossRef][ISI][Medline]
Uhlig, S., and Wollin, L. (1994). An improved setup for the isolated perfused rat lung. J. Pharmacol. Toxicol. Methods 31, 8594.[CrossRef][ISI][Medline]
Uhlig, S., and Heiny, O. (1995). Measuring the weight of the isolated perfused rat lung during negative pressure ventilation. J. Pharmacol. Toxicol. Methods 33, 147152.[CrossRef][ISI][Medline]
Vassallo, J. D., Hicks, S. M., Born, S. L., and Daston, G. P. (2004). Roles for epoxidation and detoxification of coumarin in determining species differences in clara cell toxicity. Toxicol. Sci. 82, 2633.
von Bethmann, A. N., Brasch, F., Nusing, R., Vogt, K., Volk, H. D, Müller, K. M., Wendel, A., and Uhlig, S. (1998). Hyperventilation induces release of cytokines from perfused mouse lung. Am. J. Respir. Crit. Care Med. 157, 263272.
Vodicka, P., Koskinen, M., Vodickova, L., Stetina, R., Smerak, P., Barta, I., and Hemminki, K. (2001). DNA adducts, strand breaks and micronuclei in mice exposed to styrene by inhalation. Chem. Biol. Interact. 137, 213227.[CrossRef][ISI][Medline]
West, J. A. A., Buckpitt, A. R., and Plopper, C. G. (2000). Elevated airway GSH resynthesis confers protection to Clara cells from naphthalene injury in mice made tolerant by repeated exposures. J. Pharmacol. Exp. Ther. 294, 516523.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
G. A. Csanady and J. G. Filser |













) inhaling ST at constant concentrations of up to 630 ppm (not shown: SO level of 0.30 ± 0.015 nmol/g in the rat lung exposed to 984 ± 308 ppm ST). Dashed lines: fitted by eye. Symbols ± bars: means ± SD of the values obtained by means of 